Manual New - Narishige Pc10

Over time, heating and cooling will distort the coil geometry. Inspect it weekly to ensure it remains a perfect circle centered around the capillary path.

The PC-10 features an intuitive, analogue-digital hybrid interface designed for quick adjustment. Turns the unit on or off. Mode Selector (1-Stage / 2-Stage):

Remove the finished pipettes vertically to prevent touching the ultra-fragile tips against the metal frame or heating coil. narishige pc10 manual new

Keep the unit away from air conditioning vents, windows, and heavy foot-traffic areas. Ambient air currents directly alter heater coil temperature consistency.

If the tip is too long and thin, decrease the heater setting or decrease the second-stage delay time. 6. Maintenance and Troubleshooting Cleaning the Platinum Coil Over time, heating and cooling will distort the

: Do not loosen internal structure clamps or attempt to disassemble for repair. 3. Operational Procedures Initial Setup

The Narishige PC-10 is a high-precision vertical needle puller designed for the production of glass micropipettes Turns the unit on or off

The Narishige PC-10 remains a gold standard in manual, gravity-driven pipette fabrication due to its robust design and simplistic operation. By systematically documenting your lab's specific heater settings and keeping the vertical slider completely clean, your PC-10 will deliver reproducible microelectrodes daily.

Carefully remove any transport ties securing the upper and lower slider assemblies.

Increase the Heater 1 setting or increase the Magnet strength. This keeps the glass fluid longer during the extension phase.

Typically yields "firm" pipettes suitable for patch-clamping or penetrative injections. NARISHIGE WEB NEWS Troubleshooting & Error Messages If your unit is displaying an error code, refer to the official Narishige FAQ for guidance: